Catheter Localization Coil
A catheter localization coil is the micro sensing coil mounted at the distal end of a catheter, guidewire or interventional tool. Inside an external magnetic field generator it produces a signal that the navigation system converts into position and orientation data. Chengpin builds these coils to print for overseas medical device OEMs, navigation system developers and catheter R&D teams.
Send your drawing for a DFM review.
What Is a Catheter Localization Coil?
A catheter localization coil is a miniature induction winding and the sensing element of an electromagnetic tracking system. Bonded into the distal section, it senses the alternating field produced by the external generator, and the software converts that signal into the position and orientation of the tip.
The coil is only one part of the measurement chain, so its geometry cannot be judged alone. The number of axes, the winding layout, the fixation in the shaft and the lead routing all influence the result. Final accuracy is set by the whole tracking system, the assembly process and the reconstruction algorithm, so we publish no absolute accuracy figure for a bare coil.
Two sensor families cover most projects. A 5DOF single-axis or multi-axis sensor gives position plus a limited set of orientation angles, and suits devices that constrain one rotational degree of freedom. A 6DOF multi-axis coil resolves full position and orientation where the tool can rotate freely.
Describe the degrees of freedom you need to resolve.
Coil Configurations for Catheter Navigation
These are the winding architectures we build most often. Each is a custom item defined by the drawing and the DFM review.
5DOF Single-Axis Sensor Coil
A single-axis electromagnetic tracking sensor coil is wound on a fine core or mandrel and senses the field component in one direction. Packaged at defined angles, several single-axis coils deliver a 5DOF solution while keeping each winding small and predictable.
6DOF Multi-Axis Tracking Coil
A 6DOF sensor coil stacks orthogonal windings so that all field components are captured at one point. Orthogonality, concentricity and inter-winding spacing must be held tightly, because small mechanical errors appear in the reconstructed orientation.
Distal-Tip Micro Coil
The distal-tip micro coil sits as close to the tip as the design allows. Wire down to 0.015 mm and coil outer diameters from 0.30 mm are achievable, subject to geometry confirmation.
Integrated Coil-and-Lead Assembly
Many programmes take the sensor as a semi-assembly rather than a bare winding. As a medical micro coil OEM and custom catheter coil manufacturer we can supply the wound coil bonded to its leads, with defined lead length, termination preparation and strain relief.
Send the coil detail of your drawing.
Materials and Insulation Options
Material selection follows the customer drawing and the applicable project requirements. The table lists materials we routinely process, not a statement that any material is approved for a particular device.
| Material / coating | Typical use in navigation coils |
|---|---|
| Ultra-fine enameled copper | General sensing windings where a thin build is required |
| High-temperature copper alloy | Windings that must tolerate elevated process or sterilisation temperatures |
| Polyimide coating | Thin, mechanically robust insulation on fine wire |
| PTFE coating | Low-friction insulation and wire handling in tight lumens |
| Parylene coating | Conformal thin-film insulation applied to the finished coil |
| Medical stainless / Nitinol mandrel | Support mandrels and cores, evaluated per drawing |
Biocompatibility is not automatic for any material or coating. Where required, we coordinate ISO 10993 testing on request and provide material documentation for your own biological evaluation.
Tell us the insulation on your drawing.
Typical Custom Specifications
These ranges describe what our winding equipment can hold. They are typical and customizable values, not catalogue stock; final numbers are fixed at DFM.
| Parameter | Typical / customizable range |
|---|---|
| Wire diameter | 0.015 - 0.10 mm |
| Coil outer diameter | 0.30 - 2.00 mm |
| Coil length | 2 - 30 mm |
| Winding layers | Single-layer or multi-layer |
| Axes | Single-axis or 3-axis |
| Lead length | 20 - 1000 mm |
| Inductance / resistance | Custom per drawing, verified by test |
| Tolerance | Per drawing, confirmed at DFM |
Send your dimensional drawing and we will confirm the applicable ranges.
Engineering Controls and OEM Support
- Winding tension control, monitored through the run to keep turns and build consistent.
- Outer diameter and length inspection against the drawing at defined stages.
- DCR and inductance testing, with results recorded per part or per lot as agreed.
- Lead routing and strain relief defined with the customer to suit the assembly method.
- Material lot traceability documentation available on request.
- Sample MOQ from 50 pieces for design verification.
- Target sample lead time of 3-7 working days, subject to DFM confirmation and material availability.
ISO 13485 requirements are reviewed per project and documentation is available on request. Where a requirement depends on your device and your notified body, we confirm it in writing.
Share your supplier questionnaire and we will respond point by point.
Applications
- Cardiac electrophysiology catheter navigation
- Interventional and vascular procedure navigation
- Flexible endoscopes and bronchoscopes
- Biopsy needle and ablation needle tracking
- Robotic surgical tool localization
- Research and laboratory test platforms
Projects that also need a passive component often start from the same build path as our Custom Precision Coil. Sensing catheter teams also review our Electrophysiology Catheter Coil.
Tell us which application you are developing.
FAQ
1. What is your MOQ?
Sample MOQ starts from 50 pieces; production quantities are confirmed at quotation.
2. What is the sample lead time?
Our target is 3-7 working days after drawing and material are confirmed, subject to DFM and availability.
3. Should I choose 5DOF or 6DOF?
Choose 5DOF when the device constrains one rotational degree of freedom, and 6DOF when the tool rotates freely and full orientation must be resolved.
4. How small can the coil be?
Typically wire from 0.015 mm and coil outer diameter from 0.30 mm, with lengths from 2 mm, confirmed at DFM.
5. Are the coils MRI compatible?
MRI compatibility depends on the material and the finished device. Final system compatibility must be verified by the device OEM; we supply material data to support that verification.
6. Can you support biocompatibility?
We coordinate ISO 10993 testing on request and provide material documentation. Final biocompatibility is established for the finished device by its manufacturer.
7. Can you supply a semi-assembly?
Yes. Options include a wound and cleaned coil, a coil on a mandrel, a coil bonded to leads, or a semi-assembled module.
Request a Quote
To review your project, send us: coil OD / ID / length, wire gauge and material, turns or target inductance / DCR, axis and DOF, lead type and length, annual quantity, and test requirements. We return a DFM review with a quotation.